Sylvac SA Ch. du Closalet 16 CH- 1023 Crissier
INSTRUCTIONS
FOR USE
Program version V1.1
November 1999
Vista de página 0
1 2 3 4 5 6 ... 73 74

Resumo do Conteúdo

Página 1 - INSTRUCTIONS

Sylvac SA Ch. du Closalet 16 CH- 1023 CrissierINSTRUCTIONSFOR USEProgram version V1.1November 1999

Página 2 - CONTENTS

1.5. EXAMPLES OF OPERATIONS1/ Simple measurements:- Choose the required channel: + channel number + or - Reverse measuring direction: then 2/ Combi

Página 3 - 3 RS232 D104 Switching unit

106/ Measurements with classification (2 to 6 classes)- Enter number of classes and then limits of classes using: then activates or deactivates the

Página 4 - D100S DISPLAY UNIT

1.6 FUNCTION KEYS1.6.1 CLEAR KEY- Operating in normal measuring mode: move to following channel, adjust contrast or control lifting of probes with D11

Página 5 - 1.3 FUNCTION KEYS

121.6.2 ENTER KEY- Operating in normal measuring mode: moves directly to preceding channel, adjust contrast or controls descent of probewith D110 or D

Página 6

1.6.3 CHANNEL KEY The D100S unit can command up to 64 channels. The use of D102/D108 units makes it possible to connect up to 64P5, P10, P25 or P50 pr

Página 7

14d) reversing the measuring direction:The measuring direction is indicated by the arrow preceding the Channel number. An up arrow indicates that the

Página 8 - 1.4 OPERATION

5/ press. to confirm the input or for A+/-B+/-C press once for +C or twice for -C6/ select channel number C (1-64)7/ press to confirm inputToleranc

Página 9

163/ confirm input number with Following display will appear:The channel selected by default in 3/ will appear. A number from 1 to 64 canbe entered.4/

Página 10 - 1.5. EXAMPLES OF OPERATIONS

1.6.4 RESOL KEYAllows choice of resolution displayed and printed:1/ press The following will be displayed:2/ to obtain the correct resolution, select

Página 11

181.6.6 ANALOG KEYDisplays or removes the analog scale. The analog scale features an indicating range of 100 points, each one states oneleast signific

Página 12 - 1.6 FUNCTION KEYS

CONTENTS1 D100S Display Unit Page1.1 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 13 - 1.6.2 ENTER KEY

1.6.7 TOL KEYDisplays or removes tolerance indicators:indicates a measured value smaller than the nominal dimension + negative tolerance on the extern

Página 14 - 1.6.3 CHANNEL KEY

206/ input of upper tolerance. If negative, must always be greater than the lower limit.7/ confirm by pressing The following will be displayed:8/ ent

Página 15

1.6.8 CLASS KEY Displays or removes the classification of measured values.Indicates measured value under the 1st class limit.Indicates measured value

Página 16 - Chan. to change:

221.6.9 STAT KEY Allows two types of statistical results to be displayed:Choice of display:1/ press 2/ then appears:1 = display of classical statist

Página 17 - = Input : 2

Following display:-3σ = _X - 3σ = lower regulation or intervention limit+3σ = _X + 3σ = upper regulation or intervention limit-NG = number of stored

Página 18 - 1.6.4 RESOL KEY

243) Press , the histogram of stored values will be displayed:The number of cells is determined with the formula: n = √Nbut at least 5 cells and maxi

Página 19 - 1.6.6 ANALOG KEY

1.6.10 PRESET KEYDisplays stored preset value.The external contact (e.g. foot pedal) can also be configured to preset the displayed value.Any preset v

Página 20

261.6.11 SETUP KEYAllows input of parameters for the following functions: Stat, Class, Tol, Clear or Enter, Min/Max, Set Up, Preset, Store,Print, Scan

Página 21

3.1.2/ Display hold or store. According to the store key function (Set Up then Store), the external contact enables:- display hold: as long as the foo

Página 22 - Enter number of

283.5/ Activate/deactivate sound release of keyboard and exceed tolerance limit function.3.6/ Date/time input: adjusting permanent clock: Year - Month

Página 23 - 1.6.9 STAT KEY

21.9.2 Total reset of unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .551.9.3 Loss

Página 24 - (- NG) + (+ NG)

1.6.12 MIN/MAX KEYAllows the choice of displaying minimum, maximum, difference Max - min, or mean (max + min)/2 instead of normalmeasurement. Entering

Página 25

30Example: The probe is used for measuring a camshaft. The unit displays the maximum value. The camshaft is rotatedand the displayed value is frozen o

Página 26 - 1.6.10 PRESET KEY

1.6.13 SCAN KEYStarts scanning of 1 to n channels (n being defined as indicated below), compares each measurement with its assi-gned tolerance and dis

Página 27 - 1.6.11 SETUP KEY

3213/ Enter nominal size and tolerances for all channels to be scanned (refer to Section 1.6.7).14/ Put a block gauge (master) under the probes and be

Página 28

Prints measured values of channels 1 to n through the RS232 and Centronics outputs, according to theselected Print format (refer to Section 1.6.14).St

Página 29

341.6.14 PRINT KEYPrinting of values through:- Centronics output, identical to parallel port of PC (LPT1 or 2). This output does not need to be set up

Página 30 - 1.6.12 MIN/MAX KEY

2.2/ Peripheral choice, the following menu will appear:These different choices modify the output format for the Centronics/RS232 out-puts. Remote comm

Página 31

36-NG 7+NG 6%DEF 46.428Cp 0.2986Cpk 0.2450Cm 0.2932Cmk 0.22902.2.2/ Format for 80 column printer: for A4 printers, i.e. Centronics (standard PC link c

Página 32 - 1.6.13 SCAN KEY

To print header again: press longer than two seconds.Following pages show printing examples in 80 column format. If tolerances mode is not activated,

Página 33 - 12.402

38c/ printing of statistical values with tolerances active (printing therefore of 2nd statistical display and histogram) andwith complete header. Afte

Página 34 - PIECE 3 :

D100S DISPLAY UNIT1.1 GENERAL DESCRIPTIONThe D100S unit displays the absolute displacement of the long-travel Sylvac probes P5, P10, P25, P50. Highest

Página 35 - 1.6.14 PRINT KEY

d/ printing example of _X/R control chartSYLVAC SA Piece ident. : SHAFT Drawing nbr : PM230.010.412 Work station :

Página 36 - 4 : 15 " "

402.2.3/ Output format for 40 column printer: for small printers with a paper width of approx. 80 to 120mm.Printing time per line may be entered, as s

Página 37

c/ printing example of statistical analysis and stored values (tolerances activated):SYLVAC SA Piece ident. : SHAFT Drawing nbr

Página 38

422.2.4/ Format for 15 column printer: for small portable printers, battery or accumulator powered (e.g. EDP5000 orSP1).As previously, the unit reques

Página 39

- MM = selects metric unit.- IN = selects inch unit.- RES1 = selects resolution of 0.0001 mm or 0.00001 IN.- RES2 = selects resolution of 0.001 mm or

Página 40 - X/R control chart

44- KEY0 = keyboard locked.- KEY1 = keyboard unlocked.- RST = general reset of the unit (return to initial status).- SET? = the unit transmits general

Página 41

- UP = retraction of probes using D110/D110V.- DOW = return motion of probes using D110/D110V.- OUT 1 = activates automatic output of values:- in norm

Página 42

46b/ remote command of D100S unit:10 CLS20 OPEN"COM1: 4800, E,7, 1, CS, DS, CD" AS#130 PRINT" Input command word"40 INPUT B$ enter

Página 43

1.6.15 STORE KEYThis key may have 2 functions:1/ Display hold. A reverse video appears at the top right-hand side of the screen. This is cancelled by

Página 44

48The following will appear if the 1st choice is selected.Pressing the key erases the last stored value.The key will recover erased value(s). When

Página 45

41.3 FUNCTION KEYS1.3.1 Summary of functions:: displays or removes analog scale.: direct conversion mm/inch.and : lock/unlock the mm/inch conversion.

Página 46

1.7 Calibration of the unit1.7.1 General calibrationThe D100S units are calibrated at the factory. However if a recalibration is required, proceed as

Página 47

502/ Same instruments, but with correction of 10 points (every 2.5 mm) max. error 0.7 µm:Introduction of correction: probe out to probe in.1/ Switch

Página 48 - 1.6.15 STORE KEY

1.8 REAR PANEL1.8.1 RS-232-C Input/output9 pin D-sub female connector (external view):Pin 1: Charger output 8.5 V/300 mA non-regulated (current limit

Página 49

521.8.3 Outputs15-pin D-sub female connector (external view)Pin 1: common (digital ground).Pin 2: 0 to 5V or -2.5V to + 2.5V analog output according

Página 50 - 1.7 Calibration of the unit

Notes:- if pneumatic unit is activated, classification is only possible for 2 classes.Pin 8: Common for 8 optocoupler outputs.Max voltage = 30V, max c

Página 51

541.8.5 Socket for mains chargerMay be inserted in either 5, 6 or 7.Before insertion: ensure socket polarization is at 12 o'clock.1.8.6 Socket fo

Página 52 - 1.8 REAR PANEL

1.9 IN CASE OF DIFFICULTY1.9.1 No display on screen when unit is switched on.- the accumulators may be flat. If this is the case, you should reconnec

Página 53

561.9.4 Software versionThe software version may be displayed as follows:1/ switch the unit off 2/ press 3/ keep it pressed when switching on Then

Página 54

1.10 REPLACING THE LITHIUM MODULEIf the following message appears on the screen when the unit is switched on:This means that the data back-up lithium

Página 55

581.11 REPLACING ACCUMULATORSWhen the autonomy of the D100S unit is no longer satisfactory it is necessary to replace the accumulators inside theunit

Página 56 - 1.9 IN CASE OF DIFFICULTY

5: in MIN/MAX mode: clear the MIN/MAX memory to the current value.: introduction of parameters or functions for the following keys: Start, Class, Tol,

Página 57

1.12 TECHNICAL SPECIFICATIONSEnclosure: in terblend plastic (= ASA + polycarbonate):- resistant to alcohol, glycols, most oils and greases, diluted a

Página 58 - OUT OF MEMORY

60Outputs: - analog 0 to 5V or -2.5 to +2.5V, output current: max 10 mA. Voltage error:+/- 4 % max. 0 position: +/- 100 mV.- absolute analog 6 to 10 V

Página 59 - 1.11 REPLACING ACCUMULATORS

1.14 ACCESSORIESOrder No- Sylvac SP1 printer 926.1807- Connecting cables for: - parallel printer (Centronics) 925.5630- PC AT computer (Dsub 9p cable

Página 60 - 1.12 TECHNICAL SPECIFICATIONS

622 MULTICHANNEL UNITS: 2 CHANNELS (D102) & 8 CHANNELS (D108)2.1 GENERAL DESCRIPTIONThe D102 and D108 multichannel units are designed to work with

Página 61 - 1.13 DELIVERY

2.4 TECHNICAL SPECIFICATIONSD102/D108 units: Housing in Terblend plastic (= ASA + Polycarbonate: refer to characteristicsdescribed in Section 1.12).

Página 62 - 1.14 ACCESSORIES

643 D104 4-CHANNEL RS232 SWITCHING UNIT3.1 GENERAL DESCRIPTIONThe D104 switching unit is designed to link up to four OptoRS simplex or duplex instrume

Página 63 - Common In/Out87654321

3.4 TECHNICAL SPECIFICATIONSTerblend plastic housing (= ASA+Polycarbonate: refer to characteristics inSection 1.12).Polyurethane legs with clip-on at

Página 64 - 2.5 ACCESSORIES

664 SYLVAC PROBES P5, P10, P25 AND P504.1 GENERAL DESCRIPTIONSylvac long travel probes are of compact design and are distinctive by their stability an

Página 65 - 3.3 OPERATION

4.3 USE4.3.1 Precautions- To ensure optimum measurement precision avoid all lateral pressure when presenting the probe contact to the objectto be meas

Página 66 - 3.5 ACCESSORIES

684.5 TECHNICAL SPECIFICATIONSAccuracy using extension cables:These measuring errors are applicable only when using D100S or D101 units without recali

Página 67 - 4.2 PROBE DIMENSIONS

6: either proceed to preceding channel, adjust contrast or pneumatic descentWhen entering parameters: confirm numerical entry or proceed to following

Página 68 - 4.4 MAINTENANCE

4.6 ACCESSORIESStandard measuring tip with 2 mm diameter ball (supplied with each probe)Order No: 905.2204Ordinary comparator tips, with standard M2.5

Página 69 - 4.5 TECHNICAL SPECIFICATIONS

705 PNEUMATIC COMMAND UNITS D110 AND D111 or VACUUM D110V and D111V5.1 GENERAL DESCRIPTIONThe pneumatic command units D110/D111 or D110V/D111V (vacuum

Página 70 - 4.6 ACCESSORIES

5.4 OPERATION OF D111 OR D111V UNIT1/ Configure the 1st D110/D110V unit as shown in Section 5.3.2/ Connect the air supply input marked 'In'

Página 71 - 5.3 OPERATION OF D110 UNIT

72Packaging in synthetic material includes:Order No- 1 D110 unit (16 channels) 904.1110- or 1 D110V unit (16 vacuum channels) 904.1112with:- 1 command

Página 72 - 5.5 TECHNICAL SPECIFICATIONS

Version 99.12 / SYL - D100S - E / ALINEA 99.12 /200 681.038-110

Página 73 - 5.6 ACCESSORIES

1.3.2 General method:All functions are directly accessible, e.g. by pressing the Tol key, tolerance indicators will be displayed. Pressing it againcau

Página 74

86/ Select language by pressing then then and then 1,2 or 3 (see chap. 1.6.11).This selection is memorized permanently by the unit(along with all o