- MM = selects metric unit.
- IN = selects inch unit.
- RES1 = selects resolution of 0.0001 mm or 0.00001 IN.
- RES2 = selects resolution of 0.001 mm or 0.0001 IN.
- RES3 = selects resolution of 0.01 mm or 0.001 IN.
- RES4 = selects resolution of 0.1 mm or 0.01 IN.
- ANA0 or ANA OFF = no analog display.
- ANA1 or ANA ON = displays analog scale.
- TOL0 = no display of tolerances indicators.
- TOL1 = displays tolerances indicators.
- TOL 10.2 0.1 -.05 I = input of nominal size 10.2, upper tolerance 0.1, lower tolerance -0.05 and internal
measurement (= I).
- TOL? = output of memorized tolerances values: for example 10.000 0.005 -0.003 E.
- CLA0 = no display of class indicators.
- CLA1 = displays class indicators.
- CLA3 12.1 12.2 12.3 12.4 = input of 3 classes, lower limit of class 1 is 12.1, upper limit of class 1 is 12.2, upper
limit of class 2 is 12.3 and upper limit of class 3 is 12.4 (according to current unit).
- CLA? = output of memorized class values.
- MAX = selects max function.
- MIN = selects min function.
- DEL = selects delta function (max - min).
- MEA = selects mean function (max + min/2).
- CLE = re-initializes min/max registers when unit is in max, min, delta or mean modes.
- ENT = displays in succession Max - Min - Delta - Mean... in Min/Max mode.
- NOR = re-establishes normal measuring function.
- MOD? = the unit sends its measuring mode: NOR, MAX, MIN, DEL or MEA.
- STO = when receiving this command, the unit will store the measured value of the current
channel. maximum of 6000 measurements can be stored, shared between the chan-
nels defined for storage. For example if STO = 4, we have 6000/4 = 1500 values that
can be stored per channel.
- STO4 = gives the number of channels for storage, starting with channel 1. Here, 4 channels
are reserved for storage —> 1500 measurements per channel. Reinitializes the Store
memory.
- STO0 = Removes Hold from the display.
- STO? = the unit outputs stored values for current channel, from 1st stored value to the last.
- STO1? = outputs stored values of channel 1.
- STO RST = reset complete store memory.
- STO 3 RST = reset store memory of channel 3.
- STA? = the unit outputs statistical analysis of current channel.
- STA2? = outputs statistical analysis of channel 2.
- SCA = initiates scanning.
- SCA 21 = sets number of channels to be scanned, here 21.
- SCA2 6 = initializes scanning from channel 2 to channel 6 and starts scanning.
- SCA? = the unit returns the stored number of channels to scan.
- ENT = has the same function as the Enter key in scanning mode (D100S displays next table
or quit scanning mode).
- CLE = has the same function than Clear key in scanning mode (D100S displays preceding
table or quit scanning mode).
43